User's Guide ______________________________________________________________________
36 __________________________________________________________________ M211060EN-C
Set Serial Interface Mode
SMODE [xxx]<cr>
where
xxx = STOP, RUN, or POLL
In STOP mode: outputting only when command is issued, any command
can be used.
In RUN mode: outputting automatically, only command S can be used. In
POLL mode: outputting only when command is issued, any command
can be used after the line has been opened using the OPEN command.
The polling mode can be used when more than one probe is connected to
one serial bus. When set to POLL mode, the probes communicate one at
a time when the specific address is called on the serial line. See
descriptions of the commands ADDR on page 38 and OPEN on page 45.
Examples:
smode stop
Serial mode : STOP
smode run
Serial mode : RUN
smode poll
Serial mode : POLL
NOTE
In the RUN mode, the probe may send the measurement data message
right as you are typing the S command to stop the sending. Therefore,
you may need to repeat the S command. This must be noted especially
when designing computer programs to access the probe. You can also
use the Esc key on your keyboard to stop the sending.