Chapter 7 ____________________________________________________________ Technical Data
VAISALA _______________________________________________________________________ 61
Operating Environment
Operating temperature range
HMP60 -40 ... +60 °C
HMP110 -40 ... +80 °C
HMP110T -40 ... +80 °C
Operating humidity range 0 ... 100 %RH, non-condensing
Electromagnetic compatibility EN 61326-1: Electrical equipment for
measurement, control and laboratory use -
EMC requirements – for use in industrial
locations.
EN 55022 Class B: Information technology
equipment - Radio disturbance characteristics
- Limits and methods of measurement.
Inputs and Outputs
Operating voltage
with 0...1 V / 2.5 V or RS-485 5 ... 28 V
with 0...5 V / 1 ... 5 V 8 ... 28 V
Current consumption
typical average 1 mA
max peak 5 mA
Start-up time
probes with analog output 4 s at operating voltage 13.5 ... 16.5 VDC
2 s at other valid operating voltages
probes with digital output 1 s
Outputs
analog output channels 2
analog output types 0 ... 1 VDC / 2.5 VDC / 5VDC, 1 ... 5 VDC
with loop power converter 4 ... 20 mA (separate module, compatible
with humidity accuracy only)
digital output (optional) RS-485 two-wire half-duplex
External loads
0 ... 1 R
L
min 10 k
0 ... 2.5 / 0 ... 5 / 1 ... 5 V R
L
min 50 k
Mechanics
Materials
body Stainless steel (AISI 316)
grid filter Chrome coated ABS plastic
cable Polyurethane
Housing classification IP65
Body thread M12x1 / 10 mm
Cable connector 4-pin M8 (IEC 60947-5-2)
Cable lengths 0.3 and 3 m
Weight
probe 17 g
probe with 0.3 m cable 28 g