Vaisala HMP60 Guide de l'utilisateur Page 62

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 66
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 61
User's Guide ______________________________________________________________________
60 __________________________________________________________________ M211060EN-C
Performance (HMP110)
Relative Humidity
Measurement range 0 ... 100 %RH
Accuracy (incl. non-linearity, hysteresis
and repeatability)
temperature range +0 ... +40 °C
0 ... 90 %RH ±1.7 %RH
90 ... 100 %RH ±2.5 %RH
temperature range -40 ... 0 °C
and +40 ... +80 °C
0 ... 90 %RH ±3.0 %RH
90 ... 100 %RH ±4.0 %RH
Factory calibration uncertainty (+20 °C) ±1.5 %RH
Humidity sensor Vaisala HUMICAP® 180R
Temperature
Measurement range -40 ... +80 °C
Accuracy over temperature range
0 ... +40 °C ±0.2 °C
-40 ... 0 °C, +40 ... +80 °C ±0.4 °C
Temperature sensor Pt1000 RTD 1/3 Class B IEC 751
Dewpoint
Measurement range -40 ... +80 °C
Accuracy (incl. non-linearity,
hysteresis and repeatability)
temperature range 0 ... +40 °C
when temperature - dewpoint
< 15 °C ±1 °C
15 ... 25 °C ±2 °C
temperature range -40 ... 0 °C
and +40 ... +80 °C when
temperature - dewpoint < 15 °C ±2 °C
Performance (HMP110T)
Temperature
Measurement range -40 ... +80 °C
Accuracy over temperature range
0 ... +40 °C ±0.2 °C
-40 ... 0 °C, +40 ... +80 °C ±0.4 °C
Temperature sensor Pt1000 RTD 1/3 Class B IEC 751