Vaisala BAROCAP PTB330 Guide de l'utilisateur Page 28

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 144
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 27
USER'S GUIDE____________________________________________________________________
26 __________________________________________________________________ M210855EN-D
Grounding the Cables
To ensure the best possible EMC performance, you must ground the screen of the electrical cable as shown in
1212-005
Figure 16 below and in the instructions below that.
1212-005
Figure 16 Grounding the Screen of Electrical Cable
To ground the electrical cables:
1. Cut back outer sheath to desired length.
2. Cut back screen braiding or screen foil to dimension X (see
1212-005
Figure 16 above).
3. Push the domed cap nut (item 1) and the seal insert with contact
socket of the gland (item 2+3) onto the cable as shown in the
diagram.
4. Bend over the screen braiding or screen foil by about 90° (item
4).
5. Push the seal insert with the contact socket of the gland (item
2+3) up to the screen braiding or screen foil.
Vue de la page 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 143 144