USER'S GUIDE____________________________________________________________________
70 __________________________________________________________________ M210855EN-D
Setting and testing the analog outputs
AMODE
Displays the analog output mode (if an
AOUT-1 module is connected).
ASEL
Sets the analog output quantity and scaling
(low/high).
ACAL
Adjusts the analog output.
AERR
Sets the analog output error value.
ATEST
Starts/ends an analog output test.
Setting and testing the relays
RSEL
Sets the relay scaling (if a RELAY-1
module is connected).
RTEST
Starts/ends a relay output test.
General Settings
Changing Quantities and Units
NOTE
Only the quantities selected when ordering the device can be selected
as a display output quantity.
FORM
Use the serial line command FORM to change the format or define
quantities for the output command SEND and for RUN mode.
FORM [x]
where:
x = Formatter string
Formatter string consists of quantities and modifiers. When selecting
the quantity, use the abbreviations of the quantities presented in Table
2 on page 15. The modifiers are presented in Table 12 on page 71.