OPERATING MANUAL_______________________________________________________________
44 _____________________________________________________________________ U263EN-1.4
press CL
the number starts flashing
if the interval needs to be changed, press CL again and enter the
new interval; otherwise press ENT
the unit (s, min, h) starts flashing
the unit can be changed with the arrow keys and acknowledged
with ENT
after this the programme returns to Mode menu
Others
Pressure compensation
The pressure is used for pressure compensation of the DRYCAP
sensor in order to ensure the best possible measurement accuracy. If
the process pressure differs from normal ambient pressure, the value
has to be entered in the transmitter memory. The pressure also has a
considerable effect on the ppm
v
value. The pressure to be entered is
the absolute pressure in hPa (for converting pressure units, see
Appendix 8).
Select 3UHV in the main menu and the following appears:
Pressing ENT returns the programme to the main menu without
changing the pressure reading.
If the pressure needs to be changed, press CL; enter the new
pressure with the arrow keys
NOTE
The pressure compensation takes place only with the security lock
j
umper connected. If the security lock jumper is not connected, the
pressure compensation is performed with the value 1013.25 hPa.