Vaisala DMP248 Manuel de service Page 53

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 145
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 52
CHAPTER 5________________________________________________________________ COMMANDS
VAISALA _________________________________________________________________________ 47
The 0D[G7GS parameter defines the maximum allowed change of the
dewpoint value during the pre-defined G7GS time. If the change in
dewpoint value exceeds the limit, the calibration is not started. The
auto-calibration is done only after process is stabilized. The
transmitter retries untill the calibration is succesfully complited. Note
that normally, this parameter does not have to be changed.
The G7GSWLPH defines the measuring time of the change in dewpoint
prior to the auto-calibration (see the parameter 0D[G7GS.
The Max corr. parameter defines the maximum correction (%RH) the
transmitter does during each auto-calibration cycle.
The 6HWWO WLPH defines the time the output values prior to the auto-
calibration are frozen after the calibration. The time is for sensor
temperature stabilization.
NOTE
Normally, the Max dTdp and the Max corr. parameters need not be
changed.
NOTE
Auto-calibration takes place only when the security lock jumper is
connected. With command <fst on>, the serial line printing indicates
the stages of the auto-calibration (n = no auto-calibration, w,h,s =
different phases in auto-calibration, S = settling time auto-calibration
is on).
DCAL<cr>
>dcal
Calibration...
any key to abort
> (appears when the auto-calibration
is completed in 60 - 70 seconds)
If you press any key, the calibration is interrupted. If no key is
pressed, the calibration takes place immediately. If the process is
unstable or relative humidity is over 10%, the prompt (>) appears
immediately. In that case the calibration is not performed. The
calibration is not performed again even if the maximum correction is
exceeded. (see Chapter Display/keypad commands)