APPENDIX 1 SERIAL COMMANDS__________________________________________________________
VAISALA _________________________________________________________________________ 83
Auto-calibration commands
DRYCAL Defining the frequency of auto-calibration cycles
DRYCAL<cr>
>drycal
DCAL ON
Interval min : 360 ?
Max dTdp: 2.00 ?
dTdp time s : 10
Max corr. : 0.040 ?
The interval parameter defines the frequency of auto-calibration
cycles. When the transmitter is turned on, the first auto-calibration
takes place after on hour unless the frequency has been set to less than
an hour. After the first auto-calibration, the set frequency is activated
and the auto-calibration takes place e.g. every six hours. If the setting
is changed, it becomes valid only after the next auto-calibration has
been completed. If you wish to activate it immediately, reset the
transmitter or turn it off.
The 0D[G7GS parameter defines the maximum allowed change of the
dewpoint value during the pre-defined G7GS time before auto-
calibration. If the change in dewpoint value exceeds the limit, the
calibration is not started. The auto-calibration is done only after
process is stabilized. The G7GS defines the time for 0D[ G7GS
measurement.
The Max corr. parameter defines the maximum correction (%RH) the
transmitter does during each auto-calibration cycle. The 6HWWO WLPH
defines the time the output values (measured before auto-calibration)
are frozen after the auto-calibration. The time is for sensor
temperature stabilization.
NOTE
Normally, the Max dTdp and the Max corr. parameters need not be
changed.
NOTE
Auto-calibration takes place only when the security lock jumper is
connected.