CHAPTER 7____________________________________________________________MAINTENANCE
VAISALA _________________________________________________________________________ 67
CHAPTER 7 MAINTENANCE
Reference measurements
Reference measurements are needed to verify whether the transmitter
readings are within specifications. This way the user can check if the
transmitter needs calibration or service.
The reference measurement should be made as close to the checked
sensor as possible and the readings should be read at the same time,
when possible. A re-calibrated DMP248 transmitter can be used as a
reference.
Self-diagnostics
The DMP248 transmitter goes through a self-diagnostics procedure
when the power is switched on. If the procedure does not reveal any
errors or faults, the transmitter starts operating normally. If errors or
faults are found, check first if the DRYCAP
sensor is damaged. If it
is intact, send the transmitter to Vaisala or a Vaisala representative for
repairs. The error messages the transmitter outputs are listed in
Appendix 5.
If any errors occur during operation, the error messages are output on
the local display if the transmitter displays measurements; if the
menus are used, error messages are not output. The LEDs indicate
errors at all times. During operation, however, the error messages are
not output automatically through the serial interface. If there is any
reason to doubt that there is something wrong with the transmitter, use
command ERRS: