Vaisala DRYCAP DMT340 SERIES Guide de l'utilisateur Page 102

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 191
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 101
User's Guide _______________________________________________________________________
100 __________________________________________________________________ M210704EN-G
Outputting Reading Once
Use the SEND command to output the reading once in STOP mode. The
output format depends on which parameters the transmitter can output.
SEND<cr>
Examples:
>send
Tdf= -5.8 'C H2O= 3715 ppmV x= 2.3 g/kg
>send
Tdf= -5.9 'C Td= -6.6 'C Tdfa= -5.9 'C Tda= -6.6 'C H2O=
3696 ppmV x= 2.3 g/kg RH= 13.9 %RH a= 2.7 g/m3 aNTP=
3.0 g/m3 T= 22.2 'C dT= 28.1 'C
Assign an Alias for the SEND Command
Use the SCOM command to assign a new command that works like the
SEND command. The standard SEND command of the transmitter will
always function normally whatever the SCOM definition may be.
Command names are case-insensitive.
SCOM<cr>
Example (assigning the command MEASURE as an alias for the SEND
command):
>scom
Send command : ? measure
>measure
Tdf= -5.8 'C H2O= 3715 ppmV x= 2.3 g/kg
>
You can remove the SCOM definition by pressing Esc when entering the
command name:
>scom
Send command : measure ? <esc>
>
Vue de la page 101
1 2 ... 97 98 99 100 101 102 103 104 105 106 107 ... 190 191

Commentaires sur ces manuels

Pas de commentaire