Vaisala DRYCAP DMT340 SERIES Guide de l'utilisateur Page 40

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 191
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 39
User's Guide _______________________________________________________________________
38 ___________________________________________________________________ M210704EN-G
NOTE
When re
-tightening the nut after detachment the nut must be tightened
without increased effort.
6. Clean and grease the tightening cone of the fitting screw after every
tenth detachment. Change the sealing washer every time the fitting
screw is detached. Use high-vacuum grease (for example Dow
Corning) or a similar grease.
0503-033
Figure 23 Cleaning of the Tightening Cone
The numbers refer to Figure 23 above:
1
=
Fitting screw
2
=
Sealing washer
3
=
Tightening cone
4
=
Clean cotton stick
CAUTION
In pressurized processes it is essential to tighten the supporting nuts and
screws very carefully to prevent
loosening of the probe by the action of
pressure.
NOTE
When the DMT340 is installed in a process with a pressure differing
from the selected operating pressure at the time of ordering, please enter
the pressure value of the process into the transmitter m
emory. You can
enter the value
in the following ways:
-
using the display/keypad option
-
using the buttons on the motherboard
-
using the serial line
-
using the Modbus protocol
See section
Pressure Compensation Setting on page 106. If you are
using
Modbus, see section Configuration Registers on page 185.