Vaisala DRYCAP DMT340 SERIES Guide de l'utilisateur Page 32

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 191
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 31
User's Guide _______________________________________________________________________
30 ___________________________________________________________________ M210704EN-G
Wiring
Cable Bushings
A single electrical cable with a screen and three to ten wires is
recommended for power and analog/serial connections. The cable
diameter should be 8 ... 11 mm. The number of cable bushings depends
on the transmitter options. See the following recommendations for the
cable bushings:
0503-010
Figure 15 Cable Bushings
The numbers refer to Figure 15 above:
1
=
Cable for signal/powering Ø8 ... 11 mm
2
=
Cable for optional module Ø8 ... 11 mm
3
=
Cable for optional power module Ø8 ... 11 mm
NOTE
When there is high electric noise level (for example near powerful
electric motor) in the operating environment it is recommended to use
shielded cable or take care that the signal cables are separated from other
cables.