User's Guide ______________________________________________________________________
34 __________________________________________________________________ M211060EN-A
View Software Version
Use the VERS command to display the software version of the probe.
VERS<cr>
Example:
vers
HMP60 / 1.00.0
Serial Line Output Commands
Start Measurement Output
Use the R command to start the continuous outputting of measurement
values as an ASCII text string to the serial line. The output always
includes readings for temperature, RH and Td.
R<cr>
Example:
r
T= 22.6 'C RH= 22.8 %RH Td= 0.3 'C
T= 22.6 'C RH= 22.5 %RH Td= 0.2 'C
T= 22.6 'C RH= 22.5 %RH Td= 0.2 'C
Outputting the results continues in intervals issued with the command
INTV. You can stop the output by entering the S command.
Stop Measurement Output
Use the S command to stop the continuous measurement output: