User's Guide ______________________________________________________________________
40 __________________________________________________________________ M211060EN-A
Example: one-point calibration (LiCl reference, 11 %RH):
crh 11
OK
Example: one-point calibration with NaCl reference (75 %RH):
crh 75
OK
Giving the command without parameters starts the two-point calibration.
Remember to allow the humidity to stabilize for 20 - 40 minutes after
changing the reference.
Example: two-point calibration with LiCl (11 %RH) and NaCL (75
%RH) references:
crh
RH : 11.2684 1. ref ? 11
Press any key when ready ...
RH : 75.0612 2. ref ? 75
OK
Clear Adjustment of RH Measurement
Use the CRHCLR command to clear the adjustment of RH measurement
that has been done using the CRH command.
CRHCLR<cr>
Example:
crhclr
OK
Calibrate Temperature Measurement
Use the CT command to perform a one-point or two-point temperature
(T) calibration. One-point calibration adjusts the offset for the
measurement, two-point calibration adjusts offset and gain.
CT [reference]<cr>
When performing a one-point calibration, you need to place the probe in
a single temperature reference and wait for 20 – 40 minutes for the
temperature to stabilize. To apply the adjustment, enter the CT command
with the reference temperature as a parameter.