Repair Maintenance
Changing the INTERCAP Sensor
(HMP60)
This procedure restores the humidity measurement accuracy of the
HMP60. No adjustment after the sensor change is needed.
To perform this procedure, you need a new INTERCAP sensor. It is also
recommended that you replace the filter with a new one. For order
codes, see section Options and Accessories on page 60.
1. Remove the filter from the probe by turning it counter-clockwise.
2. Pull out the old INTERCAP sensor (the larger of the two sensors
inside the probe) and insert a new one. Refer to Figure 14 on
page 52. Handle the new sensor by the plastic socket. DO NOT
TOUCH THE SENSOR PLATE.
3. Attach a new filter on the probe.
0912-097
Figure 14 Changing the Sensor