Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 41
Example: one-point calibration
ct 23.5
OK
Giving the command without parameters starts the two-point calibration.
Remember to allow the temperature to stabilize for 20 - 40 minutes after
changing the reference. The first reference point must be smaller than the
second point, and the difference between the reference points must be
more than 30 ºC. To update the measured value while the command is
running, press enter without inputting a value.
Example: two-point calibration
ct
T : 22.03 Ref1 ? 22
Press any key when ready ...
T : 55.12 Ref2 ? 55
OK
Clear Adjustment of T Measurement
Use the CTCLR command to clear the adjustment of temperature
measurement that has been done using the CT command.
CTCLR<cr>
Example:
ctclr
OK
View User Adjustment Parameters
Use the L command to view the current user adjustment parameters. This
command is useful for checking the currently applied customer
calibration.
L<cr>
The output values are as follows:
- Cp offset and gain: capacitance, calibrated using the CRH command
- T offset and gain: calibrated using the CT command
Example (shows default values, no user calibration done):
l
Cp offset : 0.00000000E+00
Cp gain : 1.00000000E+00
T offset : 0.00000000E+00
T gain : 1.00000000E+00