User's Guide _______________________________________________________________________
38 ____________________________________________________________________M210498en-A
24. Press BACK, EXIT, EXIT to return to the basic display.
Checking and adjusting the MMT318 transmitters
You can check and adjust MMT318 transmitter reading against a
calibrated reference probe of MMP78 or against another reference
condition by using MI70 indicator in communication. Follow the first 4
steps and continue according to the chosen calibration method.
1. Connect the DRW216050-calibration cable between the screw
terminal connector of the MMT318 transmitter and the MI70
indicator.
2. Turn on MM70.
- In case you use a MMP78 probe as a reference: The reading of the
transmitter is shown on the first or middle row of the display,
depending on the connector port to which the calibration cable is
connected. Value of the port I is shown on the upper row of the
display and the value of the port II on the lower row of the display.
Difference of the readings is shown in the lowest row.
- In case MI70 is used as an display, the reading of the MMT318
transmitter is shown on a display.
3. Remove the MMT318 transmitter unit from the mounting plate and
press the adjustment button once. Now MI70 is starting the Adjustment
mode.
4. Press OK to start adjustment.
Field checking and adjustment by
using a calibrated reference probe
Follow the steps from 1 to 4 and continue as follows:
5. Select a
w
, RH ,T or H
2
O (H
2
O can be checked only) press SELECT.