User's Guide _______________________________________________________________________
42 ____________________________________________________________________M210498en-A
CHAPTER 12
CALIBRATION AND ADJUSTMENT OF
MM70 SERIES PROBES
General about calibrations
The MM70 is fully calibrated as shipped from factory. Adjustment shall be
done if there is a reason to believe that device is not within the accuracy
specifications. The recommended calibration interval is one yar.
It is recommended to send the device to Vaisala Service Centres for
calibration and adjustment, see contact information on page 50.
Alternatively, user can calibrate and adjust MM70 by following the
instructions given in this chapter.
Preparations before the calibration
Before the calibration the used sensor should be cleaned with instrument
air to blow out existing oil or gently first flush with hepthane (C7H16) and
dry with instrument air. This is to decrease response time and to prevent
contamination of the salt bath and change of the reference condition. See
page 48.
NOTE
It is important to clean the sensor before calibration as the dirty
sensor can contamine the salt bath and change the reference
condition.