Chapter 12 ________________________________ Calibration and adjustment of MM70 series probes
VAISALA _______________________________________________________________________ 45
1-point adjustment
Usually it is recommended to make an adjustment in two reference
humidities. If adjustment is done by using one reference humidity (1-point
adjustment), select the reference humidity so that it represents the
measuring environment. Make the adjustment in one point as instructed in
the previous section Two-point relative humidity adjustment (select 1-point
adjustment
and follow the display instructions).
Temperature adjustment
Temperature adjustment in one or two points can be done if there is reason
to believe that the adjustment is changed. If adjustment is done only in one
point, reference condition shall represent the measuring environment.
1. Take out the screw covering the calibration button (in the probe
handle).
2. Press the calibration button. For pressing, use tool with thin and
sharp point, such as small screwdriver. When pressing the button,
indicator turns to adjusting mode.
3. Select T, press SELECT.
4. Now the adjustment mode is on, press ADJUST to select the
adjustment method: 1-point adjustment or 2-point adjustment.
1-point adjustment
Follow the steps from 1 to 4 and continue as follows:
5. Select 1-point adjustment, press SELECT.
6. Set the probe to a reference temperature. You can follow the
stabilization from the GRAPH display. Press READY when the
reading is stabilized in the reference.
7. Give the reference temperature value by using the arrow buttons.
Press OK.
8. Confirm the adjustment, press YES. (By pressing NO you return
to adjustment mode display and no changes are made).
9. Calibration is carried out. Press BACK to exit the adjustment mode
and EXIT to return to the basic display.