2-point adjustment by using a
calibrator
If you use humidity calibrator HMK15, please use the adapter fitting (13.5
mm) on the measurement hole.
Follow the steps from 1 to 4 and continue as follows:
5. Select the quantity press SELECT.
6. Remove the filter from the transmitter's probe and insert the probe
head into the first reference condition.
7. Press ADJUST to continue adjusting.
8. Select 2-point adjustment, press SELECT.
9. Press READY when the reading is stabilized in the first reference
condition (takes about 30 minutes or more).
10. Enter correct reference value by using the arrow buttons. Press
OK.
11. Remove the probe and insert the probe head into the second
reference condition. You can follow the stabilization from the
GRAPH display (takes about 30 minutes or more).
12. Press READY when the reading is stabilized in the second reference
condition.
13. Give the higher reference humidity value by using the arrow buttons.
Press OK.
14. Confirm the adjustment, press YES (by pressing NO you return
to adjustment mode display and no changes are made). If the
difference between two references is less than 50 %, adjustment can
not be done.
15. Adjustment is done. Press
BACK to exit the adjustment mode press
EXIT to return to the basic display.
16. Detach the calibration cable.