Vaisala HMP260 SERIES Manuel de service Page 35

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 118
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 34
HMP260 SERIES
HMP260-U017en-1.5 Operating Manual
1997-11-25 29
5.3.6.1 Setting the serial interface operation mode
Select Mode in the main menu; the following is displayed:
Select Serial output:
The currently valid setting flashes. Select the desired mode with the
arrow keys and press ENT. After this the programme returns to the
Mode Menu.
When Run mode is selected, the currently valid output interval is dis-
played:
The output interval setting can be changed as follows:
press CL
the number starts flashing
if the interval needs to be changed, press CL again and enter the new
interval; otherwise press ENT
the unit (seconds or hours) starts flashing
the unit can be changed with the arrow keys and acknowledged with
ENT
after this the programme returns to Mode menu
5.3.7 Others
5.3.7.1 Setting the measurement integration time
By lengthening the measurement integration time any stray changes in the out-
put can be filtered out: the transmitter calculates the average of a number of
measurement cycles defined by the user.
Vue de la page 34
1 2 ... 30 31 32 33 34 35 36 37 38 39 40 ... 117 118

Commentaires sur ces manuels

Pas de commentaire