HMP260 SERIES
HMP260-U017en-1.5 Appendix 1: Serial commands
FDATE Adding date to output
FDATE xxx <cr>
xxx = ON or OFF
When FDATE is activated, the current date is output at the beginning of the output line.
The time is set with command DATE. After RESET or power on the current date is
1991-01-01.
Activating the date output
>fdate on
Form. date : ON
>r
1995-03-10 RH= 21.1 %RH T= 26.0 'C
1995-03-10 RH= 21.3 %RH T= 26.0 'C
1995-03-10 RH= 23.1 %RH T= 26.0 'C
1995-03-10 RH= 22.2 %RH T= 26.0 'C
1995-03-10 RH= 20.6 %RH T= 26.0 'C
...
Inactivating the date output
>fdate off
Form. date : OFF
>r
RH= 20.2 %RH T= 26.0 'C
RH= 19.9 %RH T= 26.0 'C
RH= 19.8 %RH T= 26.0 'C
RH= 19.7 %RH T= 26.0 'C
RH= 19.7 %RH T= 26.0 'C
...
SERI Serial bus settings
SERI b p d s x <cr>
b = bauds (300, 600, 1200, 2400, 4800, 9600)
p = parity (n = none, e = even, o = odd)
d = data bits (7 or 8)
s = stop bits (1 or 2)
x = duplex (H = half, F = full)
Giving the command on its own outputs the current settings:
SERI <cr>
4800 E 7 1 FDX