User's Guide ______________________________________________________________________
100 _________________________________________________________________ M210855EN-C
0705-211
Figure 43 Current/Voltage Switches of Output Modules
The example below illustrates how to select a 0 ... 5 V voltage output.
0706-031
Figure 44 Dip Switch Selection Example
The number refers to Figure 43 on page 100:
1 = Current/voltage selection output switches (from 1 to 2)
2 = Current/voltage range selection switches (from 3 to 7) in
analog output 1 and 2.
3 = Switch for service use only. Keep in OFF position always.
NOTE
Only one of the switches 1 or 2, must be ON at a time. Only one of the
switches 3 to 7, must be ON at a time.