Chapter 6 __________________________________________________ Calibration and Adjustment
VAISALA______________________________________________________________________ 115
Feeding Adjustment Information
This information is shown on the device information fields, see section
Information Display on page 57.
Using Display/Keypad
1. If you are not in the adjustment menu, press the ADJ button on the
motherboard to open the ADJUSTMENT MENU.
2. Select Adjustment info, press the right arrow button.
3. Select Date, press the SET button. Enter date by using the arrow
buttons. Press the OK button.
4. Select i, press the SET button. Enter information text including 17
characters at maximum. Use the arrow buttons. Press the OK
button.
5. Press the EXIT button to return to the basic display.
Using Serial Line
CTEXT
Use the CTEXT command to enter text to the adjustment information
field.
Example:
>ctext Vaisala/MSL
Calibration text : Vaisala/MSL
>
CDATE
Use the CDATE command to enter date to adjustment information field.
Set the adjustment date in format YYYY-MM-DD.
>cdate 2006-06-12
Calibration date : 2006-06-12
>