User's Guide ______________________________________________________________________
74 __________________________________________________________________ M210855EN-C
VERS Displays the product name and software
version number.
SNUM Shows the device and module serial
numbers.
Measurement commands
R Changes the serial mode to RUN and
starts outputting measurement results
according to the FORM string (with
interval defined by INTV).
INTV [0...255 s/min/h/d]Shows or sets the continuous output
interval (for RUN mode).
SEND [0...99] Shows the measurement results
according to the configured form.
SMODE [STOP/POLL/
RUN]
Shows or sets the start mode.
SDELAY [0...254] (0... 2540
ms)
Shows or sets the answer delay for the
serial line in tens of milliseconds.
ADDR [0...255] Sets the barometer address.
OPEN [0...99] Opens communications after having
entered the CLOSE command.
CLOSE Closes communications until the OPEN
command is entered.
SCOM Shows or sets an alias (a user-specific
form) for the SEND command. The
given alias cannot be a command already
in use.
SEND Shows the measurement results
according to the configured form.
Measurement setting commands
TQFE [-40...+60°C] Shows or sets the temperature for QFE
corrected pressure.
DPMAX Shows or sets the maximum difference
pressure between barometer modules.
HHCP [ -30...+30 m] Shows or sets the altitude for height
corrected pressure.