Chapter 4 _________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 101
The AutoCal is carried out if the following criteria for the
measurement environment are fulfilled:
- Relative humidity must be <10 %.
- Temperature must be 0<T<140 ºC.
- Humidity environment must be stable. The maximum change in the
dewpoint can be 2 ºC in 15 seconds.
NOTE
AutoCal cannot operate if the above conditions are not fulfilled.
If the adjustment in the AutoCal reaches a preset maximum value or if
the AutoCal correction fails, for example, because of unstable
conditions, a new AutoCal will take place later (if the automatic
AutoCal is turned on).
Automatic AutoCal
As a default, the automatic AutoCal in DMT345 and DMT346 is
turned on. In this mode, the calibration takes place automatically if the
dewpoint or temperature changes significantly, typically more than
10 ºC. However, if there are no changes in the conditions, the AutoCal
will take place repeatedly after one hour from the last AutoCal.
Manual AutoCal
To ensure that AutoCal has taken place for obtaining the most
accurate measurement in a very dry environment, you can perform
Autocal manually before measuring as follows:
1. Open the MAIN MENU by pressing any of the ▲▼◄► arrow
buttons.
2. Select Measuring, press the arrow button.
3. Select Manual AutoCal, press the arrow button.
4. Press START to start AutoCal. If the AutoCal conditions are
not fulfilled, a note appears on the display informing that the
calibration cannot be made.