
User's Guide _______________________________________________________________________
122 __________________________________________________________________ M210762EN-A
Error States
In error state the quantity is not measured and the output is shown as
follows:
- analog channel outputs 0 mA or 0 V (you can use the serial line
command AERR or display/keypad to change this fault indication
value, see section Analog Output Fault Indication Setting on page
90)
- the serial port outputs stars (***)
- the cover LED is blinking
- optional display: error indicator is lit.
0506-055
Figure 53 Error Indicator and Error Message
Number refers to Figure 53 above:
1 = Error Indicator
- The error indicator disappears when the error state is over and you
have checked the error message. Press the INFO button to display
the error message.
Table 28 Error Messages
Error Message Action
Humidity sensor measurement
malfunction.
Check the integrity of the humidity
probe and the probe cable. Clean
the probe from dirt, water, ice or
other contaminants.
Humidity sensor short circuit Check the integrity of the humidity
probe and the probe cable. Clean
the probe from dirt, water, ice or
other contaminants.
Humidity sensor open circuit Check the integrity of the humidity
probe and the probe cable.
Temperature sensor open circuit. Check the integrity of the humidity
probe and the probe cable.
Temperature sensor short circuit. Check the integrity of the humidity
Commentaires sur ces manuels