Vaisala DRYCAP DMT346 Guide de l'utilisateur Page 90

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 146
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 89
User's Guide _______________________________________________________________________
88 ___________________________________________________________________ M210762EN-A
1. Open the MAIN MENU by pressing any of the ▲▼◄► arrow
buttons.
2. Select Interfaces by pressing the arrow button.
3. Select Analog outputs by pressing the arrow button.
4. Select Output 1/2/3 by pressing the arrow button.
5. Select Quantity by pressing the ▲▼ arrow buttons. Confirm
your selection by pressing CHANGE.
6. Select the quantity by using the arrow buttons. Press SELECT
to confirm your selection.
7. Select Scale, lower limit, by pressing the ▲▼arrow buttons.
Press SET to confirm your selection. Press OK to confirm your
setting.
8. Select the upper limit by pressing the ▲▼arrow buttons. Use
the arrow buttons to set the upper limit value. Press SET to
confirm your selection. Press OK to confirm your setting.
9. Press EXIT to return to the basic display.
AMODE/ASEL
Use the serial line to select and scale the analog output quantities.
Connect the transmitter to the PC. Open the terminal connection
between your PC and the transmitter.
1. Check the analog outputs with the AMODE command.
Example:
>amode
Ch1 output : 0 ... 1V
Ch2 output : 0 ... 1V
>
2. Select and scale the quantities for the analog outputs with the
command ASEL [xxx yyy zzz]. Note that the optional quantities
can be selected only if they have been selected when ordering
the device.
where
xxx = Quantity of channel 1
y
y
y
= Quantity of channel 2
zzz = Quantity of the optional analog output channel 3