Chapter 5 ___________________________________________________ Calibration and Adjustment
VAISALA_______________________________________________________________________ 117
Adjusting Temperature Using Serial
Line
CT
Use the serial line command CT to adjust the temperature.
1. Press the ADJ button on the motherboard to open the
adjustment mode.
2. Remove the probe filter and insert the probe head into the
reference temperature.
3. Type CT and press Enter. Type C and press Enter to check if
the reading is stabilized.
4. Let the reading stabilize, enter the reference temperature after
the question mark and press Enter three times.
When having another reference temperature (two-point
adjustment) press Enter twice and insert the probe to the second
reference. When the reading is stabilized, enter the second
reference temperature after the question mark and press Enter.
Please, note that the difference between the two temperature
references must be at least 30 ºC.
Example (1-point adjustment):
>ct
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? 16.0
Press any key when ready ...
T : 16.06 Ref2 ?
OK
>
OK indicates that the calibration has succeeded.
5. Press the ADJ button on the motherboard to close the
adjustment mode.
6. Take the probe out of the reference conditions and replace the
filter.