Chapter 4 _________________________________________________________________ Operation
VAISALA________________________________________________________________________ 53
CHAPTER 4
OPERATION
This chapter contains information that is needed to operate this
product.
Getting Started
Within a few seconds after power-up the led on the cover of the
transmitter is lit continuously indicating normal operation. When the
transmitter is turned on the first time, the language selection window
opens: Select the language with ▲▼ arrow buttons and press the
SELECT button.
The pressure has an effect on humidity calculations and accuracy.
Therefore, accurate calculations can be achieved only when the
ambient pressure is taken into consideration. See section Pressure
Compensation Setting
on page 73 for instructions on how to set the
pressure.
Start-up time for DMT345/346 transmitter is in total about 6 minutes.
The outputs (serial and analog) are activated 3 seconds after powering
up the transmitter. In addition, 10 seconds after the measurement the
outputs will freeze for about 6 minutes due to the sensor self
diagnostics procedure (Sensor Purge and AutoCal). The frozen output
value will be the value the transmitter reached during the 10 seconds
of measurement. After the self diagnostics procedure the outputs are
operational again.