Vaisala DRYCAP DMT346 Guide de l'utilisateur Page 40

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 146
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 39
User's Guide _______________________________________________________________________
38 ___________________________________________________________________ M210762EN-A
4. Bend over the screen braiding or screen foil by about 90º
(item 4).
5. Push the seal insert with the contact socket of the gland
(items 2 & 3) up to the screen braiding or screen foil.
6. Mount lower part (item 5) on the housing
7. Push the seal with the contact socket of the gland and
(item 2 & 3) flush into the lower part (item 5).
8. Screw the domed cap nut (item 1) onto the lower part (item 5).
Transmitter Housing Grounding
In case you need to ground the transmitter housing, the grounding
connector is found inside the housing (see Figure 1 on page 18) Note
anyhow that the probe head is connected to the same potential as the
housing. Make sure that different groundings are made to the same
potential. Otherwise harmful ground currents may be generated.
If it is needed to have galvanic isolation of the power supply line from
the output signals, DMT346 can be ordered with optional output
isolation module. This module prevents harmful grounding loops.