User’s Guide ______________________________________________________________________
166 _________________________________________________________________ M210906EN-C
General Unit Settings
0XF!0XF,f=11111111&11100010,o=AAC1DB1A,c=A263,
i=HEL___,n=A3430012,2=2528,3=3512 <cr><lf>
Wind Configuration Settings
0WF!0WF,g=A,l=N,t=A,0=273.00,1=273.01,2=273.00,3=273.00,4=2
73.00,5=273.00,a=45.1,b=50.2,u=54.9,v=63.1,x=65.1,y=65.1<cr><l
f>
Table 23 General Unit Settings
Field Character Field Name Description
f Factory options Selection of parameters
o Order code Ordering identity as delivered
(10 characters)
c Calibration date Y=2003, A, B,…=2005, 2006,
1..52 = week, 1...7, weekday
i Info Factory signature (10
characters)
n Device s/n A,B,...=2005,2006..., 1..52 =
week,
1...7 = weekday , 1...9999 =
serial number
2 2.5 V reference 2500mV (default)
3 3.5 V reference 3500mV (default)
Table 24 Wind Configuration Settings
Field Character Field Name Description
g Strategy A=All, N=North, E=East,
S=South
l Pulse length N=Normal,auto, A=Adjusted on
half, S=Short, E=Extended,
T=Test
t Single transducer mode A=All, N=North, E=East,
S=South
0..5 Zero adjustment 1...655.35 us (default 273.00 us)
a,b Detect level btw. N and E 0...100 % (default 70 %)
u,v Detect level btw. E and S 0...100 % (default 70 %)
x,y Detect level btw. S and N 0...100 % (default 70 %)