User's Guide _______________________________________________________________________
104 __________________________________________________________________ M210762EN-A
0504-051
Figure 49 Activating Start-up Purge
1. Press ON to activate Start-up Purge. Pressing OFF disables the
function.
2. Press EXIT to return to the basic display.
PUR
With the PUR command you can enable or disable interval and
power-up Purge and set the interval for interval purge. The default
purge interval is 24 hours.
It is not recommended to change these settings.
Type PUR and press Enter to proceed.
Example:
>pur
Interval Purge : ON ?
Interval : 1440 min ?
Power-up Purge : ON ?
Duration : 60 s ?
Settling : 240 s ?
Temperature : 180 'C ?
Temp. diff. : 0.5 'C ?
Trigger Purge : OFF ?
RH trigger : 20 %RH ?
>
NOTE
When Purge in power up is enabled, wait about 5 min after power up
before taking measurements. The output channels are locked for the
first operation minutes to the initial measured values.