User's Guide _______________________________________________________________________
94 ___________________________________________________________________ M210762EN-A
0605-040
Figure 46 Relay Availability
Arrow in Figure 46 above shows where enabled relays are listed on
the display. Activation state shown in black. Disabled relays are not
shown.
Use the display/keypad to set the relay outputs.
1. Open the MAIN MENU by pressing any of the ▲▼◄► arrow
buttons.
2. Select Interfaces, confirm by pressing the arrow button.
3. Select Relay outputs, confirm by pressing the arrow button.
4. Select Relay 1/2/3/4, confirm by pressing the arrow button.
5. Select the Quantity, confirm by pressing Change. Select the
Quantity by using the arrow buttons. Confirm your selection by
pressing Select. (Press Fault Status when the relay follows the
transmitter error.) Press Change to set the value.
6. Select Act. above / Act. below. Press SET to confirm your
selection. If asked, select MODIFY if you want to set the
setpoint by using the arrow buttons. Select REMOVE if you
want to remove the setpoint.
7. Select Hysteresis Press SET. Set the hysteresis by using the
arrow buttons. Press OK.
8. Select Relay enable, press ON/OFF to enable/disable the relay.