USER'S GUIDE____________________________________________________________________
106 _________________________________________________________________ M210704EN-E
INTV
Use the command INTV[xxx yyy] to set the outputting interval for the
RUN mode.
where
xxx
= Output inverval (0 ... 255). 0: the fastest possible output rate.
yy
= Unit (s, min or h)
Example:
>INTV 10 min
Output intrv. : 10 min
ECHO
Use the command ECHO [x] to set the user port echo. The command
either enables or disables echo of characters received.
ECHO [x]
where
x = ON (default) or
= OFF
NOTE
You can use the SERI, SMODE, INTV and ECHO commands to
change/view the user port settings even if you are currently connected to
the service port.
Data Filtering
The averaging data filter calculates a average over a certain period of
time. The lowest measurement noise is achieved with the extended
filtering. There are three filtering levels available.
Table 29 Filtering Levels
Setting Filtering level
OFF No filtering
ON (default) Standard = short filtering (about 15 s moving average)
EXTENDED Extended filtering (default: 1 min average)
Use display/keypad to set the filtering level.