USER'S GUIDE____________________________________________________________________
144 _________________________________________________________________ M210704EN-E
Adjusting Dewpoint T
d/f
Before adjusting dewpoint, first carry out the two-point relative humidity
adjustment, which will ensure the basic adjustment level. After that,
continue with the actual T
d/f
adjustment.
Two-point Relative Humidity Adjustment
using Display/Keypad
For DMT340 M-sensor, the humidity references of 0 % (for example
Nitrogen) and 10 ... 20 % are required.
NOTE
For the DMT340 M-sensor both reference humidities must be below
20% RH.
To make the adjustment, follow the instructions below. Use preferably
the display keypad and let the display information guide you through the
two-point RH adjustment process.
1. Press the Adjustment button ADJ on the motherboard inside the
transmitter to enable adjustments.
2. Select Adjust Td measurement, press the ►arrow button.
3. Select 2-point RH adjustment, press Start. Press OK to perform
purge of point 1.
0706-018
Figure 74 Starting the Adjustment
0706-019
Figure 75 Performing Purge