Chapter 5 ______________________________________________________________ Maintenance
VAISALA ______________________________________________________________________ 137
Table 30 Error Messages
Error Code Error Message Action
0 Humidity sensor
measurement malfunction.
Check the integrity of the humidity probe and the
probe cable. Clean the probe from dirt, water,
ice or other contaminants.
1 Humidity sensor short
circuit
Check the integrity of the humidity probe and the
probe cable. Clean the probe from dirt, water,
ice or other contaminants.
2 Humidity sensor open
circuit
Check the integrity of the humidity probe and the
probe cable.
3 Temperature sensor open
circuit.
Check the integrity of the humidity probe and the
probe cable.
4 Temperature sensor short
circuit.
Check the integrity of the humidity probe and the
probe cable. Clean the probe from dirt water, ice
or other contaminants.
5 Temperature
measurement malfunction
Check the integrity of the humidity probe and the
probe cable. Clean the probe from dirt water, ice
or other contaminants.
6 Temperature sensor
current leak.
Check the integrity of the humidity probe and the
probe cables. Clean the probes from dirt, water,
ice or other contaminants.
7 Internal ADC read error Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.
9 Checksum error in the
internal configuration
memory
Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.
10 Internal EEPROM read
error
Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.
11 Internal EEPROM write
error
Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.
12 … 13 Add-on module 1 (or 2)
connection failure
Turn off the power and check the module
connection. Turn on the power.
14 Device internal
temperature out of range
Ensure that the operating temperature is within
the valid range.
15 Operating voltage out of
range
Ensure that the operating voltage is within the
valid range.
18 Internal ADC reference
voltage out of range
Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.
19 Internal analog output
reference voltage out of
range
Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.
20 … 23 Configuration switches for
analog output 1/2/3 set
incorrectly
Check and re-set the switches, see Analog
Output Settings on p
age 114.
24 … 25 EEPROM failure on add-
on module 1 (or 2)
Disconnect the power and check the analog
output module connection.
26 Communication module
installed in incorrect add-
on module slot
Disconnect the power and change the
communication module to another module slot.
28 … 29 Unknown/incompatible
module installed in add-on
module slot 1(or 2)
Ensure that the module is compatible with the
DMT340.