USER'S GUIDE____________________________________________________________________
134 _________________________________________________________________ M210704EN-E
Example:
>purge
Purge started, press any key to abort.
>
The prompt '>' appears when the heating period is over. However, the
transmitter outputs are locked to values measured before purge until the
settling time is over.
Sensor Warming
DMT340 contains a sensor warming function, that is, when humidity
reaches a set limit (default: 70%RH), sensor warming operation is
activated automatically. Sensor warming continues as long as the
humidity is above that limit.
When the warming starts, the sensor operation indicator is displayed on
the optional display and relays configured as "online status" (if any) will
be released. Warming does not affect T
d/f
,T
d/f atm,
T
d
,T
d atm,
x, H2O and Pw
quantities, all others will be frozen.