Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 129
where
In STOP mode: measurements output only by command, all commands
can be used
In RUN mode: outputting automatically, only command S can be used
to stop, see command INTV below.
In POLL mode: measurements output only with command SEND. See
command ADDR below.
When several transmitters are connected to the same line, each
transmitter must be entered an own address in the initial configuration,
and POLL mode must be used.
INTV
Use the INTV [n xxx] command to set the RUN mode output interval.
where
n
= 1 - 255
xxx
= S, MIN or H
Sets the RUN mode output interval. The time interval is used only when
the RUN mode is active. For example, the output interval is set to 10
minutes.
>INTV 10 min
Output intrv. : 10 min
>
Setting RUN output interval to zero enables the fastest possible output
rate.
ADDR
Addresses are required only for POLL mode (see serial line command
SMODE on page 105). Use the ADDR [aa] command to input the RS-
485 transmitter address.
where
aa
= address (0 ... 99) (default = 0)
Example: the transmitter is entered the address 99.
>ADDR <cr>
Address : 2 ? 99