USER'S GUIDE____________________________________________________________________
130 _________________________________________________________________ M210704EN-E
OPEN [nn]
When all transmitters on the RS-485 bus are in POLL mode the OPEN
[nn] command sets one transmitter temporarily to STOP mode so that
other commands can be entered.
where
nn = address of the transmitter (0 ... 99)
CLOSE
The CLOSE command switches the transmitter back to the POLL mode.
Example:
>OPEN 2 (opens the line to transmitter 2, other
transmitters stay in POLL mode)
>CRH(for example, calibration performed)
...
>CLOSE (line closed)
Sensor Functions
AutoCal
To obtain the best possible accuracy in measurements taken in dry
environments, DMT340 has a built-in AutoCal. During the AutoCal,
DMT340 adjusts the dry-end reading to correspond to the calibrated
values. This is a unique and patented method to avoid errors in accuracy
when monitoring low dewpoints.
The AutoCal is carried out if the following criteria for the measurement
environment are fulfilled:
- Relative humidity must be <2 % (DMT340 M-sensor).
- Temperature must be 0<T<80 ºC (or sensor warming activated for -
40<T<80 ºC).
- Humidity environment must be stable. The maximum change in the
dewpoint can be 2 ºC in 15 seconds.
NOTE
AutoCal cannot operate if the above conditions are not fulfilled.