BGC-FTIR | Data Acquisition 35
Analog and Switched Voltage
The Temperature/Humidity Sensor KPC 1/6 ME and the Pyranometer CMP3
transmit analog voltage which is measured at the DIFF-Ports of the Datalogger
CR1000.
Figure 1.23: Connecting to SE- and DIFF-Ports
Lambrechts Wind Speed 14557 transmits analog current. Anyway, this signal
is also measured at the DIFF-Port of the Datalogger CR1000 but has to be
transformed into analog voltage by using a high-precision shunt resistor of 100
Ohm. The Temperature Probe T107 has to be excited by the EX-Port because
of switched voltage and is measured at the SE-Port. For details see section
1.3.3 on page 42 and 45.
Serial Data
The digital pressure sensor PTB210 transmits serial data which are received
at the COM-Port of the Datalogger CR1000. For connection see section 1.4.6
on page 45 . For setting up like shown in section 1.2.6 on page 25 simply
connect to the RS232-Interface on your PC or Laptop according to figure 1.24
and start communication with a terminal program.
Figure 1.24: Connecting to COM-Ports and RS232
NOTE: Take care to use the right parameters for communication through
the serial line. (section 1.2.6 on page 25)