BGC-FTIR | Programming 41
The powerup.ini is always copied to the CF-Card with the associated files
for proceeding the action. See section A.3 on page 52 for the default file.
At power-up this file will copy the program on the CF-Card to the CPU of the
datalogger for running immediately. Data on the CF-Card will be preserved.
The new data will be simply added to the old tables if there are no changes in
saving of data. Otherwise the data will be stored in new tables.
NOTE: Do not use editors like WordPad, because it will attach header in-
formations to the file causing it to abort. Check the file with the
CR1000KD to see what the CR1000 actually sees.