However, regression shows linear characteristics towards the line-up of booth sen-
sors. Apart from that accuracy amoung the two sensors in temperature reading is
well demonstrated, but is to be discussed later in this section.
Figure 5.4 shows the humidity profile of both sensors regarding the reference device
of the climate chamber.
Figure 5.4: RH Profile Sensor 1 and Sensor 2
Both sensors indicate quite linear charactericstics and there are just small variances
among the sensors in bands less and greater than 70 % humidity. However, the
actual result shows that readings range within the specified measuring tolerance of
2 % humidity.