Vaisala WXT520 Guide de l'utilisateur Page 88

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 171
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 87
User’s Guide ______________________________________________________________________
86 __________________________________________________________________ M210906EN-C
Command format: aCx!
The response:
atttnn<cr><lf>
Start Concurrent Measurement with
CRC (aCC)
Command format: aCCx!
This command has the same function as aC but a three-character CRC
is added to the response data strings before <cr><lf>.
where
a = Device address
C = Start concurrent measurement command
x = The desired measurement
1 = Wind
2 = Temperature, humidity and pressure
3 = Precipitation
5 = Supervisor
If x is left out, the query refers to combined data
message in which the user can request data from
several sensors with just one command. See the
examples below.
! = Command terminator
where
a = Device address
ttt = The measurement completing time in seconds
nn = The number of the measured parameters available
(maximum number is 20)
<cr><lf> = Response terminator
NOTE
For changing the message parameters, units and other settings, see
Chapter 8, Sensor and Data Message Settings, on page 107.
Vue de la page 87
1 2 ... 83 84 85 86 87 88 89 90 91 92 93 ... 170 171

Commentaires sur ces manuels

Pas de commentaire